Testing Methods Validation of Copper in Bottled Drinking Water Using Flame Atomic Absorption Spectrophotometry

Sa'adah Eva, Ari Surya Winata

Abstract


Testing Laboratories of Institute for Industrial Research and Standardization (IIRS) Pontianak has used modified SNI 01-3554-2006 clause 2.22.2.1 in the testing of copper in bottled drinking water. Modification of the methods required validation. Validation aims to evaluate the suitability of such methods by means of laboratory study as standard testing methods in Various Commodity Testing Laboratory of IIRS Pontianak. Validation parameter conducted namely accuracy, precision, recovery, linearity methods, method detection limit, quantization limit, sensitivity, and measurement uncertainty. Based on the results of the validation methods and statistical data processing it was concluded that copper testing method in bottled drinking water using Flame Atomic Absorption Spectrophotometry (AAS) was compliant and suitable for use as standard testing methods in Various Commodity Testing Laboratory of IIRS Pontianak, indicated with good value of accuracy and precision, recovery 102.61%, method linearity correlation coefficient 0.9997, LOD 0.001 mg / L, LOQ 0.002 mg / L, sensitivity 0.0002 and uncertainty 0.0134 mg / L.

Keywords


AAS; copper; bottled drinking water; validation methods

Full Text:

PDF


DOI: http://dx.doi.org/10.36974/jbi.v1i2.748

Refbacks

  • There are currently no refbacks.


BIOPROPAL Industri

Published by :

Institute for Industrial Research and Standardization (Baristand Industri) in Pontianak

Agency for Industrial Research and Development, Ministry of Industry 

Jl. Budi Utomo No. 41 Pontianak, West Kalimantan, Indonesia

Tel / Fax : +62 561 881393, 881533

email      : biopropal.industri@gmail.com

 

BIOPROPAL Industri indexed in: 

Hasil gambar untuk gambar doajHasil gambar untuk gambar google scholar